A Coaxiality Measurement Method for Hole Using Non-Diffracting Beam Technology

X. B. Zhang, X. Q. Jiang, Z. Li

Research output: Contribution to journalArticlepeer-review

7 Citations (Scopus)

Abstract

Combining the non-diffracting beam datum-line method with the roundness measuring technology, a high precision coaxiality measurement method for a hole whose diameter is more than 40mm is proposed. The principle of the measuring system is introduced and error analysis and test to verify it are presented. The measuring uncertainty of this coaxiality system is less than 1μm when the measuring range is within 1.5m. The measuring uncertainty of error of roundness is less than 0.4μm when the error measuring range is -200∼+200μm. This system has many advantages and it can be applied widely for coaxiality measurement.

Original languageEnglish
Pages (from-to)283-288
Number of pages6
JournalKey Engineering Materials
Volume295-296
DOIs
Publication statusPublished - 15 Oct 2005
Externally publishedYes

Fingerprint

Dive into the research topics of 'A Coaxiality Measurement Method for Hole Using Non-Diffracting Beam Technology'. Together they form a unique fingerprint.

Cite this