An in vitro investigation into the cutting action of ultrasonic radicular access preparation instruments

M. Waplington, P. J. Lumley, L. Blunt

Research output: Contribution to journalArticle

10 Citations (Scopus)

Abstract

The aim of this study was to investigate the dentine cutting action of a CT4 and SJ4 radicular preparation inserts (Excellence in Endodontics/Analytic Technology, Orange, CA, USA) used in a Piezon Master 400 ultrasonic handpiece (Electro Medical Systems SA, Nyon, Switzerland). Calibration of the ultrasonic inserts was carried out at minimum, medium and maximum power settings using light microscopic measurement of the tip displacement amplitudes (X120 mag). Polished dentine samples were instrumented, at monitored loads, at the three chosen power settings, to produce uniform dentine cuts. The resultant dentine debris was collected for SEM analysis, and comparison with that produced by rotary instruments. The depth of dentine cut was measured at X30 magnification. Fracture of the SJ4 insert was noted at medium power setting. Results revealed a significant increase in both displacement amplitude and depth of dentine cut with power setting (ANOVA P<0.001). Dentine chip size increased with power, and differed in structure and appearance from those produced by rotary instruments. In conclusion, ultrasonic cutting is influenced significantly by power setting, and dentine chip formation is fundamentally different to rotary instrumentation.

Original languageEnglish
Pages (from-to)158-161
Number of pages4
JournalDental Traumatology
Volume16
Issue number4
DOIs
Publication statusPublished - 2000
Externally publishedYes

Fingerprint

Dentin
Ultrasonics
In Vitro Techniques
Endodontics
Switzerland
Calibration
Analysis of Variance
Technology
Light

Cite this

@article{50c2aa88616545db85f4258371968151,
title = "An in vitro investigation into the cutting action of ultrasonic radicular access preparation instruments",
abstract = "The aim of this study was to investigate the dentine cutting action of a CT4 and SJ4 radicular preparation inserts (Excellence in Endodontics/Analytic Technology, Orange, CA, USA) used in a Piezon Master 400 ultrasonic handpiece (Electro Medical Systems SA, Nyon, Switzerland). Calibration of the ultrasonic inserts was carried out at minimum, medium and maximum power settings using light microscopic measurement of the tip displacement amplitudes (X120 mag). Polished dentine samples were instrumented, at monitored loads, at the three chosen power settings, to produce uniform dentine cuts. The resultant dentine debris was collected for SEM analysis, and comparison with that produced by rotary instruments. The depth of dentine cut was measured at X30 magnification. Fracture of the SJ4 insert was noted at medium power setting. Results revealed a significant increase in both displacement amplitude and depth of dentine cut with power setting (ANOVA P<0.001). Dentine chip size increased with power, and differed in structure and appearance from those produced by rotary instruments. In conclusion, ultrasonic cutting is influenced significantly by power setting, and dentine chip formation is fundamentally different to rotary instrumentation.",
keywords = "Apicectomy, Retrograde cavity, Ultrasonic instrumentation",
author = "M. Waplington and Lumley, {P. J.} and L. Blunt",
year = "2000",
doi = "10.1034/j.1600-9657.2000.016004158.x",
language = "English",
volume = "16",
pages = "158--161",
journal = "Dental Traumatology",
issn = "1600-4469",
publisher = "Blackwell Munksgaard",
number = "4",

}

An in vitro investigation into the cutting action of ultrasonic radicular access preparation instruments. / Waplington, M.; Lumley, P. J.; Blunt, L.

In: Dental Traumatology, Vol. 16, No. 4, 2000, p. 158-161.

Research output: Contribution to journalArticle

TY - JOUR

T1 - An in vitro investigation into the cutting action of ultrasonic radicular access preparation instruments

AU - Waplington, M.

AU - Lumley, P. J.

AU - Blunt, L.

PY - 2000

Y1 - 2000

N2 - The aim of this study was to investigate the dentine cutting action of a CT4 and SJ4 radicular preparation inserts (Excellence in Endodontics/Analytic Technology, Orange, CA, USA) used in a Piezon Master 400 ultrasonic handpiece (Electro Medical Systems SA, Nyon, Switzerland). Calibration of the ultrasonic inserts was carried out at minimum, medium and maximum power settings using light microscopic measurement of the tip displacement amplitudes (X120 mag). Polished dentine samples were instrumented, at monitored loads, at the three chosen power settings, to produce uniform dentine cuts. The resultant dentine debris was collected for SEM analysis, and comparison with that produced by rotary instruments. The depth of dentine cut was measured at X30 magnification. Fracture of the SJ4 insert was noted at medium power setting. Results revealed a significant increase in both displacement amplitude and depth of dentine cut with power setting (ANOVA P<0.001). Dentine chip size increased with power, and differed in structure and appearance from those produced by rotary instruments. In conclusion, ultrasonic cutting is influenced significantly by power setting, and dentine chip formation is fundamentally different to rotary instrumentation.

AB - The aim of this study was to investigate the dentine cutting action of a CT4 and SJ4 radicular preparation inserts (Excellence in Endodontics/Analytic Technology, Orange, CA, USA) used in a Piezon Master 400 ultrasonic handpiece (Electro Medical Systems SA, Nyon, Switzerland). Calibration of the ultrasonic inserts was carried out at minimum, medium and maximum power settings using light microscopic measurement of the tip displacement amplitudes (X120 mag). Polished dentine samples were instrumented, at monitored loads, at the three chosen power settings, to produce uniform dentine cuts. The resultant dentine debris was collected for SEM analysis, and comparison with that produced by rotary instruments. The depth of dentine cut was measured at X30 magnification. Fracture of the SJ4 insert was noted at medium power setting. Results revealed a significant increase in both displacement amplitude and depth of dentine cut with power setting (ANOVA P<0.001). Dentine chip size increased with power, and differed in structure and appearance from those produced by rotary instruments. In conclusion, ultrasonic cutting is influenced significantly by power setting, and dentine chip formation is fundamentally different to rotary instrumentation.

KW - Apicectomy

KW - Retrograde cavity

KW - Ultrasonic instrumentation

UR - http://www.scopus.com/inward/record.url?scp=0034252939&partnerID=8YFLogxK

U2 - 10.1034/j.1600-9657.2000.016004158.x

DO - 10.1034/j.1600-9657.2000.016004158.x

M3 - Article

VL - 16

SP - 158

EP - 161

JO - Dental Traumatology

JF - Dental Traumatology

SN - 1600-4469

IS - 4

ER -