Assessingthe material loss of the modulartaper interface in retrieved metal-on-metal hip replacements

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

Measuring the amount of material loss in the case of revised hip replacements is considered to be a prerequisite of understanding and assessing the true in vivo performance of the implant. This paper outlines a method developed by the authors for quantifying taper material loss as well as more general taper interface parameters. Previous studies have mostly relied on visual inspection to assess the material loss at the taper interface, whereas this method aims to characterize any surface and form changes through the use of an out-of-roundness measurement machine. Along with assessing the volumetric wear, maximum linear penetration and taper contact length can also be determined. The method was applied to retrieved large head metal-on-metal femoral heads in order to quantify the material loss at this junction. Material loss from the female femoral head taper can be characterized as a localized area that is in contact with the stem taper surface. The study showed that this method has good repeatability and a low level of interoperability variation between operators.

LanguageEnglish
Article number025002
JournalSurface Topography: Metrology and Properties
Volume3
Issue number2
DOIs
Publication statusPublished - 1 Apr 2015

Fingerprint

tapering
Metals
metals
interoperability
Interoperability
stems
Inspection
Wear of materials
inspection
penetration
operators

Cite this

@article{23b7ed2fb81b4f76955aef39de9ac6a3,
title = "Assessingthe material loss of the modulartaper interface in retrieved metal-on-metal hip replacements",
abstract = "Measuring the amount of material loss in the case of revised hip replacements is considered to be a prerequisite of understanding and assessing the true in vivo performance of the implant. This paper outlines a method developed by the authors for quantifying taper material loss as well as more general taper interface parameters. Previous studies have mostly relied on visual inspection to assess the material loss at the taper interface, whereas this method aims to characterize any surface and form changes through the use of an out-of-roundness measurement machine. Along with assessing the volumetric wear, maximum linear penetration and taper contact length can also be determined. The method was applied to retrieved large head metal-on-metal femoral heads in order to quantify the material loss at this junction. Material loss from the female femoral head taper can be characterized as a localized area that is in contact with the stem taper surface. The study showed that this method has good repeatability and a low level of interoperability variation between operators.",
keywords = "Hip, Stylus, Taper, Wear",
author = "Bills, {Paul J.} and R. Racasan and P. Tessier and Blunt, {L. A.}",
year = "2015",
month = "4",
day = "1",
doi = "10.1088/2051-672X/3/2/025002",
language = "English",
volume = "3",
journal = "Surface Topography: Metrology and Properties",
issn = "2051-672X",
publisher = "IOP Publishing Ltd.",
number = "2",

}

TY - JOUR

T1 - Assessingthe material loss of the modulartaper interface in retrieved metal-on-metal hip replacements

AU - Bills, Paul J.

AU - Racasan, R.

AU - Tessier, P.

AU - Blunt, L. A.

PY - 2015/4/1

Y1 - 2015/4/1

N2 - Measuring the amount of material loss in the case of revised hip replacements is considered to be a prerequisite of understanding and assessing the true in vivo performance of the implant. This paper outlines a method developed by the authors for quantifying taper material loss as well as more general taper interface parameters. Previous studies have mostly relied on visual inspection to assess the material loss at the taper interface, whereas this method aims to characterize any surface and form changes through the use of an out-of-roundness measurement machine. Along with assessing the volumetric wear, maximum linear penetration and taper contact length can also be determined. The method was applied to retrieved large head metal-on-metal femoral heads in order to quantify the material loss at this junction. Material loss from the female femoral head taper can be characterized as a localized area that is in contact with the stem taper surface. The study showed that this method has good repeatability and a low level of interoperability variation between operators.

AB - Measuring the amount of material loss in the case of revised hip replacements is considered to be a prerequisite of understanding and assessing the true in vivo performance of the implant. This paper outlines a method developed by the authors for quantifying taper material loss as well as more general taper interface parameters. Previous studies have mostly relied on visual inspection to assess the material loss at the taper interface, whereas this method aims to characterize any surface and form changes through the use of an out-of-roundness measurement machine. Along with assessing the volumetric wear, maximum linear penetration and taper contact length can also be determined. The method was applied to retrieved large head metal-on-metal femoral heads in order to quantify the material loss at this junction. Material loss from the female femoral head taper can be characterized as a localized area that is in contact with the stem taper surface. The study showed that this method has good repeatability and a low level of interoperability variation between operators.

KW - Hip

KW - Stylus

KW - Taper

KW - Wear

UR - http://www.scopus.com/inward/record.url?scp=84979196356&partnerID=8YFLogxK