Changes in the morphology of microgrooved stem tapers with differing assembly conditions

Research output: Contribution to journalArticle

Abstract

Background

Fretting and corrosion at the head-stem taper junction has been cited as a potential clinical concern. Material loss measurement is a vital tool for quantifying changes due to in vitro, in vivo, and/or ex-vivo implant experience. Material loss measurement requires reconstruction of pre-implantation geometry to delineate damage. This is straightforward in principle for plain machined tapers, integrating between a fitted interpolated cone and the measured data. Mathematical filtration methods have been developed to remove this texture and facilitate measurement. The extent to which the assembly process could influence filtration accuracy is not currently known.
Methods

An engagement/ disengagement study was performed on 27 head/ stem pairs using 2, 4, and 8kN impact loads. Impact was delivered in three locations; axial, 10o anterior, and 10o anteroproximal. Pull-off force was measured, and stems were measured pre and post assembly using a Talyrond 365 roundness measurement machine.
Results

An increase in the plastic deformation of microgrooves and pull-off force was noted with increasing assembly loads. Off-axis impaction resulted in reduced pull-off strength and reduced uniform microgroove distortion. Volumetric change between pre and post assembly was below the noise threshold on stem and head surfaces.
Discussion

The measurement method was shown to be capable of capturing linear microgroove deformation of 1 μm peak – peak and volume loss above 0.1mm3. Lower peak load was noted when impacting a seated head when compared with engagement of discrete heads and stems, under the same assembly conditions.
LanguageEnglish
Article number100096
Pages1-11
Number of pages11
JournalBiotribology
Volume18
Early online date25 Apr 2019
DOIs
Publication statusPublished - Jun 2019

Fingerprint

Cones
Plastic deformation
Textures
Corrosion
Geometry

Cite this

@article{174b17e39ad64f2698b85c642879ff78,
title = "Changes in the morphology of microgrooved stem tapers with differing assembly conditions",
abstract = "BackgroundFretting and corrosion at the head-stem taper junction has been cited as a potential clinical concern. Material loss measurement is a vital tool for quantifying changes due to in vitro, in vivo, and/or ex-vivo implant experience. Material loss measurement requires reconstruction of pre-implantation geometry to delineate damage. This is straightforward in principle for plain machined tapers, integrating between a fitted interpolated cone and the measured data. Mathematical filtration methods have been developed to remove this texture and facilitate measurement. The extent to which the assembly process could influence filtration accuracy is not currently known.MethodsAn engagement/ disengagement study was performed on 27 head/ stem pairs using 2, 4, and 8kN impact loads. Impact was delivered in three locations; axial, 10o anterior, and 10o anteroproximal. Pull-off force was measured, and stems were measured pre and post assembly using a Talyrond 365 roundness measurement machine.ResultsAn increase in the plastic deformation of microgrooves and pull-off force was noted with increasing assembly loads. Off-axis impaction resulted in reduced pull-off strength and reduced uniform microgroove distortion. Volumetric change between pre and post assembly was below the noise threshold on stem and head surfaces.DiscussionThe measurement method was shown to be capable of capturing linear microgroove deformation of 1 μm peak – peak and volume loss above 0.1mm3. Lower peak load was noted when impacting a seated head when compared with engagement of discrete heads and stems, under the same assembly conditions.",
keywords = "Stem taper, Taper measurement, Taper engagement, Metal-on-metal",
author = "Karl Dransfield and James Williamson and Radu Racasan and Paul Bills",
year = "2019",
month = "6",
doi = "10.1016/j.biotri.2019.100096",
language = "English",
volume = "18",
pages = "1--11",
journal = "Biotribology",
issn = "2352-5738",
publisher = "Elsevier Limited",

}

Changes in the morphology of microgrooved stem tapers with differing assembly conditions. / Dransfield, Karl; Williamson, James; Racasan, Radu; Bills, Paul.

In: Biotribology, Vol. 18, 100096, 06.2019, p. 1-11.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Changes in the morphology of microgrooved stem tapers with differing assembly conditions

AU - Dransfield, Karl

AU - Williamson, James

AU - Racasan, Radu

AU - Bills, Paul

PY - 2019/6

Y1 - 2019/6

N2 - BackgroundFretting and corrosion at the head-stem taper junction has been cited as a potential clinical concern. Material loss measurement is a vital tool for quantifying changes due to in vitro, in vivo, and/or ex-vivo implant experience. Material loss measurement requires reconstruction of pre-implantation geometry to delineate damage. This is straightforward in principle for plain machined tapers, integrating between a fitted interpolated cone and the measured data. Mathematical filtration methods have been developed to remove this texture and facilitate measurement. The extent to which the assembly process could influence filtration accuracy is not currently known.MethodsAn engagement/ disengagement study was performed on 27 head/ stem pairs using 2, 4, and 8kN impact loads. Impact was delivered in three locations; axial, 10o anterior, and 10o anteroproximal. Pull-off force was measured, and stems were measured pre and post assembly using a Talyrond 365 roundness measurement machine.ResultsAn increase in the plastic deformation of microgrooves and pull-off force was noted with increasing assembly loads. Off-axis impaction resulted in reduced pull-off strength and reduced uniform microgroove distortion. Volumetric change between pre and post assembly was below the noise threshold on stem and head surfaces.DiscussionThe measurement method was shown to be capable of capturing linear microgroove deformation of 1 μm peak – peak and volume loss above 0.1mm3. Lower peak load was noted when impacting a seated head when compared with engagement of discrete heads and stems, under the same assembly conditions.

AB - BackgroundFretting and corrosion at the head-stem taper junction has been cited as a potential clinical concern. Material loss measurement is a vital tool for quantifying changes due to in vitro, in vivo, and/or ex-vivo implant experience. Material loss measurement requires reconstruction of pre-implantation geometry to delineate damage. This is straightforward in principle for plain machined tapers, integrating between a fitted interpolated cone and the measured data. Mathematical filtration methods have been developed to remove this texture and facilitate measurement. The extent to which the assembly process could influence filtration accuracy is not currently known.MethodsAn engagement/ disengagement study was performed on 27 head/ stem pairs using 2, 4, and 8kN impact loads. Impact was delivered in three locations; axial, 10o anterior, and 10o anteroproximal. Pull-off force was measured, and stems were measured pre and post assembly using a Talyrond 365 roundness measurement machine.ResultsAn increase in the plastic deformation of microgrooves and pull-off force was noted with increasing assembly loads. Off-axis impaction resulted in reduced pull-off strength and reduced uniform microgroove distortion. Volumetric change between pre and post assembly was below the noise threshold on stem and head surfaces.DiscussionThe measurement method was shown to be capable of capturing linear microgroove deformation of 1 μm peak – peak and volume loss above 0.1mm3. Lower peak load was noted when impacting a seated head when compared with engagement of discrete heads and stems, under the same assembly conditions.

KW - Stem taper

KW - Taper measurement

KW - Taper engagement

KW - Metal-on-metal

UR - http://www.scopus.com/inward/record.url?scp=85064803618&partnerID=8YFLogxK

U2 - 10.1016/j.biotri.2019.100096

DO - 10.1016/j.biotri.2019.100096

M3 - Article

VL - 18

SP - 1

EP - 11

JO - Biotribology

T2 - Biotribology

JF - Biotribology

SN - 2352-5738

M1 - 100096

ER -