High-accuracy wavelength-change measurement system based on a Wollaston interferometer, incorporating a self-referencing scheme

Xiang Qiang Jiang, J. Kemp, Y. N. Ning, A. W. Palmer, K. T.V. Grattan

Research output: Contribution to journalArticlepeer-review

Abstract

A novel wavelength-difference measurement scheme with a Wollaston prism is presented. By using a suitable reference wavelength, a small variation in the signal wavelength can be converted into a relatively larger change in the modulated wavelength, as a result of the so-called fringe beating effect, resulting in enhanced measurement sensitivity by use of autocorrelation and Gaussian filtering techniques. From the results of a simulation carried out, we observed a wavelength variation of 0.01 nm over 15 nm or 0.1 nm over 60 nm for a typical pair of laser diodes with wavelengths of 785 and 810 nm, and wavelength variations of 0.5 nm over 40 nm or 1 nm over 110 nm for 671 -and 785-nm wavelengths. These results were partially verified by the experimental results obtained for which a resolution of 0.01 nm over a range of 2.5 nm for the first pair and 0.5 nm over 4 nm for the second pair of laser diodes was seen. The results have applications to the determination of wavelength variations in a wavelength-division multiplexing system or measurement of the wavelength changes induced in a range of optical sensors.

Original languageEnglish
Pages (from-to)4907-4912
Number of pages6
JournalJournal of Engineering and Applied Science
Volume36
Issue number21
DOIs
Publication statusPublished - 20 Jul 1997
Externally publishedYes

Fingerprint

Dive into the research topics of 'High-accuracy wavelength-change measurement system based on a Wollaston interferometer, incorporating a self-referencing scheme'. Together they form a unique fingerprint.

Cite this