Optimal architecture of CdZnTe detectors for photon counting, multispectral medical x-ray imaging

Comparison between simulation and experiment

Andrew K. Rossall, Marios E. Myronakis, Alexander Cherlin, Ian Radley, Mohamed Ayoub, Dimitra G. Darambara

Research output: Chapter in Book/Report/Conference proceedingConference contribution

Abstract

In this work, we combine a number of pixellated, high quality CdZnTe crystals of different configurations with a unique, fast electronic readout system with dynamic energy binning capabilities to develop a novel detection system capable of photon-counting, multispectral medical x-ray imaging. This technique offers significant advantages over conventional medical x-ray imaging by providing improved image quality through a reduction in beam hardening artefacts and better scatter rejection; enhanced soft tissue contrast, improving on current dual energy techniques by taking full advantage of the variation in mass attenuation coefficients of different tissues; in the development of K-edge imaging, identifying different high-Z contrast agents in a single scan; and in a significant reduction in radiation dose to the patient. This state-of-the-art readout system has eight dynamic energy bins, controlled by variable threshold voltages on comparator circuits with 16-bit counters. Using a novel technique of interchangeable CdZnTe dies, a wide range of detector configurations have been tested, providing the most complete experimental description of charge transfer effects within CdZnTe crystals. Pixellated CdZnTe crystals of thicknesses ranging between 1 and 3 mm and pixel sizes ranging between 100 and 400μm have been rigorously tested to determine the energy response, linearity, and stability of each crystal with an investigation into polarisation and charge sharing effects in order to provide a comprehensive comparison between simulation and experimental. Using the most advanced photon-counting, dynamic energy binning electronic readout system to date has allowed us to determine the optimum configuration for the electronic readout for medical x-ray imaging, providing essential information regarding the number of energy bins required for task specific medical imaging.

Original languageEnglish
Title of host publication2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record, NSS/MIC 2012
PublisherIEEE
Pages4085-4087
Number of pages3
ISBN (Print)9781467320306
DOIs
Publication statusPublished - 8 Jul 2013
Externally publishedYes
Event2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record, NSS/MIC 2012 - Anaheim, CA, United States
Duration: 29 Oct 20123 Nov 2012

Conference

Conference2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record, NSS/MIC 2012
CountryUnited States
CityAnaheim, CA
Period29/10/123/11/12

Fingerprint

Photons
counting
X-Rays
readout
detectors
photons
x rays
simulation
Medical Electronics
crystals
energy
comparator circuits
Diagnostic Imaging
configurations
electronics
Artifacts
Contrast Media
attenuation coefficients
Radiation
hardening

Cite this

Rossall, A. K., Myronakis, M. E., Cherlin, A., Radley, I., Ayoub, M., & Darambara, D. G. (2013). Optimal architecture of CdZnTe detectors for photon counting, multispectral medical x-ray imaging: Comparison between simulation and experiment. In 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record, NSS/MIC 2012 (pp. 4085-4087). [6551933] IEEE. https://doi.org/10.1109/NSSMIC.2012.6551933
Rossall, Andrew K. ; Myronakis, Marios E. ; Cherlin, Alexander ; Radley, Ian ; Ayoub, Mohamed ; Darambara, Dimitra G. / Optimal architecture of CdZnTe detectors for photon counting, multispectral medical x-ray imaging : Comparison between simulation and experiment. 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record, NSS/MIC 2012. IEEE, 2013. pp. 4085-4087
@inproceedings{b351d8b118824f0cac12ef46af136f50,
title = "Optimal architecture of CdZnTe detectors for photon counting, multispectral medical x-ray imaging: Comparison between simulation and experiment",
abstract = "In this work, we combine a number of pixellated, high quality CdZnTe crystals of different configurations with a unique, fast electronic readout system with dynamic energy binning capabilities to develop a novel detection system capable of photon-counting, multispectral medical x-ray imaging. This technique offers significant advantages over conventional medical x-ray imaging by providing improved image quality through a reduction in beam hardening artefacts and better scatter rejection; enhanced soft tissue contrast, improving on current dual energy techniques by taking full advantage of the variation in mass attenuation coefficients of different tissues; in the development of K-edge imaging, identifying different high-Z contrast agents in a single scan; and in a significant reduction in radiation dose to the patient. This state-of-the-art readout system has eight dynamic energy bins, controlled by variable threshold voltages on comparator circuits with 16-bit counters. Using a novel technique of interchangeable CdZnTe dies, a wide range of detector configurations have been tested, providing the most complete experimental description of charge transfer effects within CdZnTe crystals. Pixellated CdZnTe crystals of thicknesses ranging between 1 and 3 mm and pixel sizes ranging between 100 and 400μm have been rigorously tested to determine the energy response, linearity, and stability of each crystal with an investigation into polarisation and charge sharing effects in order to provide a comprehensive comparison between simulation and experimental. Using the most advanced photon-counting, dynamic energy binning electronic readout system to date has allowed us to determine the optimum configuration for the electronic readout for medical x-ray imaging, providing essential information regarding the number of energy bins required for task specific medical imaging.",
author = "Rossall, {Andrew K.} and Myronakis, {Marios E.} and Alexander Cherlin and Ian Radley and Mohamed Ayoub and Darambara, {Dimitra G.}",
year = "2013",
month = "7",
day = "8",
doi = "10.1109/NSSMIC.2012.6551933",
language = "English",
isbn = "9781467320306",
pages = "4085--4087",
booktitle = "2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record, NSS/MIC 2012",
publisher = "IEEE",

}

Rossall, AK, Myronakis, ME, Cherlin, A, Radley, I, Ayoub, M & Darambara, DG 2013, Optimal architecture of CdZnTe detectors for photon counting, multispectral medical x-ray imaging: Comparison between simulation and experiment. in 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record, NSS/MIC 2012., 6551933, IEEE, pp. 4085-4087, 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record, NSS/MIC 2012, Anaheim, CA, United States, 29/10/12. https://doi.org/10.1109/NSSMIC.2012.6551933

Optimal architecture of CdZnTe detectors for photon counting, multispectral medical x-ray imaging : Comparison between simulation and experiment. / Rossall, Andrew K.; Myronakis, Marios E.; Cherlin, Alexander; Radley, Ian; Ayoub, Mohamed; Darambara, Dimitra G.

2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record, NSS/MIC 2012. IEEE, 2013. p. 4085-4087 6551933.

Research output: Chapter in Book/Report/Conference proceedingConference contribution

TY - GEN

T1 - Optimal architecture of CdZnTe detectors for photon counting, multispectral medical x-ray imaging

T2 - Comparison between simulation and experiment

AU - Rossall, Andrew K.

AU - Myronakis, Marios E.

AU - Cherlin, Alexander

AU - Radley, Ian

AU - Ayoub, Mohamed

AU - Darambara, Dimitra G.

PY - 2013/7/8

Y1 - 2013/7/8

N2 - In this work, we combine a number of pixellated, high quality CdZnTe crystals of different configurations with a unique, fast electronic readout system with dynamic energy binning capabilities to develop a novel detection system capable of photon-counting, multispectral medical x-ray imaging. This technique offers significant advantages over conventional medical x-ray imaging by providing improved image quality through a reduction in beam hardening artefacts and better scatter rejection; enhanced soft tissue contrast, improving on current dual energy techniques by taking full advantage of the variation in mass attenuation coefficients of different tissues; in the development of K-edge imaging, identifying different high-Z contrast agents in a single scan; and in a significant reduction in radiation dose to the patient. This state-of-the-art readout system has eight dynamic energy bins, controlled by variable threshold voltages on comparator circuits with 16-bit counters. Using a novel technique of interchangeable CdZnTe dies, a wide range of detector configurations have been tested, providing the most complete experimental description of charge transfer effects within CdZnTe crystals. Pixellated CdZnTe crystals of thicknesses ranging between 1 and 3 mm and pixel sizes ranging between 100 and 400μm have been rigorously tested to determine the energy response, linearity, and stability of each crystal with an investigation into polarisation and charge sharing effects in order to provide a comprehensive comparison between simulation and experimental. Using the most advanced photon-counting, dynamic energy binning electronic readout system to date has allowed us to determine the optimum configuration for the electronic readout for medical x-ray imaging, providing essential information regarding the number of energy bins required for task specific medical imaging.

AB - In this work, we combine a number of pixellated, high quality CdZnTe crystals of different configurations with a unique, fast electronic readout system with dynamic energy binning capabilities to develop a novel detection system capable of photon-counting, multispectral medical x-ray imaging. This technique offers significant advantages over conventional medical x-ray imaging by providing improved image quality through a reduction in beam hardening artefacts and better scatter rejection; enhanced soft tissue contrast, improving on current dual energy techniques by taking full advantage of the variation in mass attenuation coefficients of different tissues; in the development of K-edge imaging, identifying different high-Z contrast agents in a single scan; and in a significant reduction in radiation dose to the patient. This state-of-the-art readout system has eight dynamic energy bins, controlled by variable threshold voltages on comparator circuits with 16-bit counters. Using a novel technique of interchangeable CdZnTe dies, a wide range of detector configurations have been tested, providing the most complete experimental description of charge transfer effects within CdZnTe crystals. Pixellated CdZnTe crystals of thicknesses ranging between 1 and 3 mm and pixel sizes ranging between 100 and 400μm have been rigorously tested to determine the energy response, linearity, and stability of each crystal with an investigation into polarisation and charge sharing effects in order to provide a comprehensive comparison between simulation and experimental. Using the most advanced photon-counting, dynamic energy binning electronic readout system to date has allowed us to determine the optimum configuration for the electronic readout for medical x-ray imaging, providing essential information regarding the number of energy bins required for task specific medical imaging.

UR - http://www.scopus.com/inward/record.url?scp=84881585124&partnerID=8YFLogxK

U2 - 10.1109/NSSMIC.2012.6551933

DO - 10.1109/NSSMIC.2012.6551933

M3 - Conference contribution

SN - 9781467320306

SP - 4085

EP - 4087

BT - 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record, NSS/MIC 2012

PB - IEEE

ER -

Rossall AK, Myronakis ME, Cherlin A, Radley I, Ayoub M, Darambara DG. Optimal architecture of CdZnTe detectors for photon counting, multispectral medical x-ray imaging: Comparison between simulation and experiment. In 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record, NSS/MIC 2012. IEEE. 2013. p. 4085-4087. 6551933 https://doi.org/10.1109/NSSMIC.2012.6551933