The Influence of Bone Cement Type on Production of Fretting Wear on the Femoral Stem Surface

A Preliminary Study

L. A. Blunt, X. Q. Jiang, L. T. Fleming, S. M. Barrans, H. Y. Zhang

Research output: Contribution to journalArticle

33 Citations (Scopus)

Abstract

Background: It has been reported that bone cement correlates with survivorship of cemented total hip replacement. However, little research has been published to investigate the influence of bone cement type on production of fretting wear on the femoral stem. Methods: In the present study, we performed six in vitro wear simulations using the same type of femoral stem (polished Exeter V40™) and three different bone cements (Simplex P, Palacos R, and CMW 3). Findings: Fretting wear was consistently reproduced on the stem surface and the wear locations compared well with the results of retrieval studies. Selected 3D surface parameters were utilised to quantitatively evaluate fretting wear and no significant difference was identified in terms of fretting wear severity between these simulations. The bone cements were all badly damaged in those sites contacting the fretting wear areas on the femoral stem. Additionally, there were plenty of wear debris present on the cement surface, and the energy dispersive X-ray analysis confirmed that it was just cement particles for Simplex P bone cement, whilst it included metallic particles for Palacos R and CMW 3 bone cements. Interpretation: This preliminary study shed some light on the influence of bone cement type on production of fretting wear on the femoral stem surface but further research is needed to gain a better understanding on this issue.

Original languageEnglish
Pages (from-to)666-672
Number of pages7
JournalClinical Biomechanics
Volume27
Issue number7
DOIs
Publication statusPublished - 1 Aug 2012

Fingerprint

Bone Cements
Thigh
Polymethyl Methacrylate
Methylmethacrylate
Hip Replacement Arthroplasties
Research
X-Rays

Cite this

@article{3502624bf3ca4a76916a0301e0922cef,
title = "The Influence of Bone Cement Type on Production of Fretting Wear on the Femoral Stem Surface: A Preliminary Study",
abstract = "Background: It has been reported that bone cement correlates with survivorship of cemented total hip replacement. However, little research has been published to investigate the influence of bone cement type on production of fretting wear on the femoral stem. Methods: In the present study, we performed six in vitro wear simulations using the same type of femoral stem (polished Exeter V40™) and three different bone cements (Simplex P, Palacos R, and CMW 3). Findings: Fretting wear was consistently reproduced on the stem surface and the wear locations compared well with the results of retrieval studies. Selected 3D surface parameters were utilised to quantitatively evaluate fretting wear and no significant difference was identified in terms of fretting wear severity between these simulations. The bone cements were all badly damaged in those sites contacting the fretting wear areas on the femoral stem. Additionally, there were plenty of wear debris present on the cement surface, and the energy dispersive X-ray analysis confirmed that it was just cement particles for Simplex P bone cement, whilst it included metallic particles for Palacos R and CMW 3 bone cements. Interpretation: This preliminary study shed some light on the influence of bone cement type on production of fretting wear on the femoral stem surface but further research is needed to gain a better understanding on this issue.",
keywords = "Bone Cement, Femoral Stem, Fretting Wear, Simulation, Total Hip Replacement",
author = "Blunt, {L. A.} and Jiang, {X. Q.} and Fleming, {L. T.} and Barrans, {S. M.} and Zhang, {H. Y.}",
year = "2012",
month = "8",
day = "1",
doi = "10.1016/j.clinbiomech.2012.02.008",
language = "English",
volume = "27",
pages = "666--672",
journal = "Clinical Biomechanics",
issn = "0268-0033",
publisher = "Elsevier Limited",
number = "7",

}

The Influence of Bone Cement Type on Production of Fretting Wear on the Femoral Stem Surface : A Preliminary Study. / Blunt, L. A.; Jiang, X. Q.; Fleming, L. T.; Barrans, S. M.; Zhang, H. Y.

In: Clinical Biomechanics, Vol. 27, No. 7, 01.08.2012, p. 666-672.

Research output: Contribution to journalArticle

TY - JOUR

T1 - The Influence of Bone Cement Type on Production of Fretting Wear on the Femoral Stem Surface

T2 - A Preliminary Study

AU - Blunt, L. A.

AU - Jiang, X. Q.

AU - Fleming, L. T.

AU - Barrans, S. M.

AU - Zhang, H. Y.

PY - 2012/8/1

Y1 - 2012/8/1

N2 - Background: It has been reported that bone cement correlates with survivorship of cemented total hip replacement. However, little research has been published to investigate the influence of bone cement type on production of fretting wear on the femoral stem. Methods: In the present study, we performed six in vitro wear simulations using the same type of femoral stem (polished Exeter V40™) and three different bone cements (Simplex P, Palacos R, and CMW 3). Findings: Fretting wear was consistently reproduced on the stem surface and the wear locations compared well with the results of retrieval studies. Selected 3D surface parameters were utilised to quantitatively evaluate fretting wear and no significant difference was identified in terms of fretting wear severity between these simulations. The bone cements were all badly damaged in those sites contacting the fretting wear areas on the femoral stem. Additionally, there were plenty of wear debris present on the cement surface, and the energy dispersive X-ray analysis confirmed that it was just cement particles for Simplex P bone cement, whilst it included metallic particles for Palacos R and CMW 3 bone cements. Interpretation: This preliminary study shed some light on the influence of bone cement type on production of fretting wear on the femoral stem surface but further research is needed to gain a better understanding on this issue.

AB - Background: It has been reported that bone cement correlates with survivorship of cemented total hip replacement. However, little research has been published to investigate the influence of bone cement type on production of fretting wear on the femoral stem. Methods: In the present study, we performed six in vitro wear simulations using the same type of femoral stem (polished Exeter V40™) and three different bone cements (Simplex P, Palacos R, and CMW 3). Findings: Fretting wear was consistently reproduced on the stem surface and the wear locations compared well with the results of retrieval studies. Selected 3D surface parameters were utilised to quantitatively evaluate fretting wear and no significant difference was identified in terms of fretting wear severity between these simulations. The bone cements were all badly damaged in those sites contacting the fretting wear areas on the femoral stem. Additionally, there were plenty of wear debris present on the cement surface, and the energy dispersive X-ray analysis confirmed that it was just cement particles for Simplex P bone cement, whilst it included metallic particles for Palacos R and CMW 3 bone cements. Interpretation: This preliminary study shed some light on the influence of bone cement type on production of fretting wear on the femoral stem surface but further research is needed to gain a better understanding on this issue.

KW - Bone Cement

KW - Femoral Stem

KW - Fretting Wear

KW - Simulation

KW - Total Hip Replacement

UR - http://www.scopus.com/inward/record.url?scp=84863612376&partnerID=8YFLogxK

U2 - 10.1016/j.clinbiomech.2012.02.008

DO - 10.1016/j.clinbiomech.2012.02.008

M3 - Article

VL - 27

SP - 666

EP - 672

JO - Clinical Biomechanics

JF - Clinical Biomechanics

SN - 0268-0033

IS - 7

ER -